3) fuses, Replacement parts, For a1463ef – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 370: For a1598ef

Advertising
background image

<6. Maintenance>

6-17

IM 11B08A01-01E

(5) Install a new HMI unit by following steps (2) to (3) in the reverse order.
(6) Plug the three cable connectors back in their original positions.
(7) Close and lock the door of the electronics section.

<Replacement Parts>

K9802QA (for

TIIS)

K9802QB

(except for the TIIS specifi cation)

K9802QC

(for Y purge system)

Screw

F0614.ai

HMI Unit Cover

HMI Unit Cover

Screws (8 places)

Figure 6.14

(3) Fuses

As shown in Figure 6.15, there are two types of fuses in the electronics section. The ratings for
both fuses are listed in Table 6.1.

Table 6.1

Fuse of electronics section

Part Number

Rated Voltage

Rated Amperage Fusing Characteristics

Notes

A1463EF

250 VAC

6.3 A

Time-delay

A1598EF

30 A

Normal

Follow the procedures below to replace.

For A1463EF

(1) Stop the operation and the supply of power.
(2) Confi rm that the LED (Green) of “POWER” is OFF.
(3) Open the door of the electronics section, using a key included in the accessory kit.
(4) Remove the fuse holder cap by rotating it counterclockwise.
(5) Remove the fuse from the cap and install a new one.
(6) Push and rotate a cap equipped with a fuse clockwise to install it in the fuse holder.
(7) Close and lock the electronics section door.

For A1598EF

(1) Stop the operation and the supply of power.
(2) Confi rm that the LED (Green) of “POWER” is OFF.
(3) Open the door of the electronics section, using a key included in the accessory kit.
(4) Replace the fuse. Fuse can be pulled out. For Y purge type, the locking bracket must be

removed. After replacing the fuse, attach the locking bracket.

(5) Close and lock the electronics section door.

2nd Edition : May 11, 2012-00