Alarm historical record – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 268

Advertising
background image

<5. EtherLCD>

5-17

IM 11B08A01-01E

Alarm Historical Record

To display the alarm history, press F2 (History).

1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5

Alarm Historical Record

Total 1 GCM #

0

Date Time ALM # Contents
11/11 14:10 31 Card ID error

10

Menu

Auto

Clean

GCM#

F1

F2

F3

F4

F5

F6

Figure 5.19

Example of the alarm historical record

F1 (Alarm):

Displays the Alarm Status screen.

F3 (Auto):

Starts to refresh the alarm status automatically at an interval of 1 second.

F5 (Clean):

Flushes the alarm history buffer and starts the Auto refresh. For this operation,
the user level C or higher is required.

F6 (GCM #): Specifi es a GCM number.

TIP

• The newer data are shown in upper rows in order.
• Auto refresh is executed as long as the top data is the newest upon the occurrence of an alarm incident.
• Between the Date Time and ALM #, the alarm status is shown as 1 (On) or 0 (Off).

Display Item

Name

Set Lower Limit Upper Limit

Unit

Remarks

Total

Total number of

alarms in the history

data table

GCM #

GCM number

0

6

0: Alarm for the entire unit

Date Time

Date Time: Month

Date Time: Day

Date Time: Hour

Date Time: Minute

1

1

00

00

12

31

23

59

Month

Day

Hour

Minute

ALM #

Alarm number

1

600

Refer to “7.1.4

Alarm

Contents”

Contents

Alphanumeric: 12 characters

* The fi rst 12 characters are

shown when the length

exceeds the limit.

Alarm information 1

0

0xFFFF

Alarm information 2

0

0xFFFF

2nd Edition : May 11, 2012-00