<Appendix D Action of External Input and Output Signals>

App.D-7

IM 11B08A01-01E

F0503.ai

*1

2s

4s

2s

4s

2s

4s

2s

4s

*1

*1

*1

Stream 1

End of Analysis

Stream 2

End of Analysis

Stream 1

End of Analysis

Stream 2

End of Analysis

AO1

AO2

AO3

AO4

DO1

(Stream

identification

flag)

DO2

(Stream

identification

signal)

DO3

(Stream

identification

signal)

Analog Hold Output

Contact Output

End of Analysis: Time when the final peak detection is completed or peak detection stop time.
*1:

One to two seconds (depending on analysis specifications)

End of Analysis:

End of peak detection time in SYS method

*1:

One to two seconds (depending on analysis specifi cations)

Figure 8 Example of Actions of Analog Output

(2) Chromatogram Output

This is a function to output chromatograms as A/O.
Up to eight outputs are available. One detector number (DET1-1, 1-2, 2-1, 2-2, 3-1, or 3-2) is

set for one A/O. The same detector number can be assigned to multiple A/Os. The auto gain

and auto zero can be set for each detector. Settings are specifi ed the detector signal setting

screen and the individual peak setting screen. These settings relate not only to the display of A/O

chromatograms but also to that of HMI and ASET chromatograms.

Data Display Cycle

The A/O chromatogram data display cycle varies depending on the sample rate specifi ed for the

detector signal.

Sample rate (ms)

A/O chromatogram data display cycle (ms)

40

80 (A chromatogram data is extracted every 80 ms from the main unit.)

80

80

160

160

*

The HMI chromatogram data display cycle is the same as above.

2nd Edition : May 11, 2012-00

App.D