17 a/o setup – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 337

Advertising
background image

<5. EtherLCD>

5-86

IM 11B08A01-01E

Display Item

Name

Set Lower Limit Upper Limit

Unit

Remarks

Slot #

1

5

Limited

DI #

1

8

Limited

Signal name

Alphanumeric: 8 characters

Process

assign

0: No process, 1: Stream

sequence, 2: Stream, 3: Range

select, 4: Cal/Val, 5: Operation

mode, 6: Alarm process

GCM number

Setup possible when 0 to 6,

process assign entails Stream

sequence, Cal (Val) and

Operation mode.

* In the case of zero, applicable

to all GCM. When process

assign entails alarm, applicable

to alarms for all GCM and

equipment.

Process

sequence

1

8

Stream

number

1

31

Number of

times

0

999

0: Consecutive, 1 to 999: times

Range

assigned

Stream

1

31

Range

assigned

Peak

1

999

Range

number

1

31

Cal/Val

number

Cal 1, Cal 2, Cal 3, Cal 4, Cal 5,

Cal 6, Val 1, Val 2, Val 3, Val 4,

Val 5, Val 6

Operation

mode

0: Run, 1: Stop, 2: Pause

Alarm level

2 or 3

Alarm number

201 to 232 (in the case of alarm

level 2), 401 to 432 (in the case

of alarm level 3)

Message

Alphanumeric: 22 characters

5.4.17 A/O

Setup

When #1 Output stream is set at 99, the analyzed value of all streams belonging to GCM are
read out. In this case, the setup for #2 Output stream and #2 Output peak become invalid. This
setup can be done independently for each A/O.

1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5

A/O Setup

(1/2)

1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5

A/O Setup

(2/2)

Slot #

1 AO # 1

Slot #

1 AO # 1

>

- GCM number 1

>

- #2 Output stream * *

- AO Chromatogram

Det 1-1

#2 Output peak * * *

- #1 Output stream * *

- #2 Percent span offset * * * * *

#1 Output peak * * *

- #2 Percent zero offset

* * * * *

- #1 Percent span offset * * * * *
- #1 Percent zero offset

* * * * *

Menu

Status

Slot#

AO#

Menu

Status

Slot#

AO#

F1

F2

F3

F4

F5

F6

F1

F2

F3

F4

F5

F6

Figure 5.98

Example of A/O setup screen

2nd Edition : May 11, 2012-00