<1. Overview>

1-72

IM 11B08A01-01E

1.12.2 Manual

status

Manual operation is performed and chromatograms are displayed in the manual status.
The operations available in the manual status are as follows.

Oven heater ON/

OFF

When the oven belongs to two GCMs, set both the GCMs in the manual

status, then click the SYS tab and open the “Temperature regulator control”

window to turn on and off the heater.

LSV heater ON/OFF Set the GCM to which the LSV belongs to the manual status, then click the

SYS tab and open the “Temperature regulator control” window to turn on

and off the heater.

FPD detector heater

ON/OFF

Set the GCM to which the FPD belongs to the manual status, then click the

SYS tab and open the “Temperature regulator control” window to turn on

and off the heater.

Hydrogen limiter

ON/OFF

When the hydrogen limiter belongs to two GCMs, set both the GCMs in

the manual status, then turn on and off the limiter in the “GCM Det Control”

window or the “SYS Det Control” window.

Detector ON/OFF

Set the GCM to which the detector belongs to the manual status, then turn

on and off the detector in the “GCM Det Control” window or the “SYS Det

Control” window.

Stream valve ON/

OFF

Set the GCM to which the stream valve belongs to the manual status, then

turn on and off the valve in the “GCM VLV Control” window.

Atmospheric

pressure balancing

valve ON/OFF

Set the GCM to which the atmospheric pressure balancing valve belongs to

the manual status, then turn on and off the valve in the “GCM VLV Control”

window.

Built-in oven valves

(RV, LSV) ON/OFF

Set the GCM to which the built-in oven valves (RV, LSV) belong to the

manual status, then turn on and off the valves in the “GCM VLV Control”

window or the “SYS VLV Control” window.

The following operation modes are available in the manual status.
The process status and the manual status can be changed to each other in the stop mode.
When the equipment is switched from the process mode to the manual mode, valves and stream
valves are all turned off.

Run

• A mode to perform measurement

• Activate this mode in the manual status to start to display a

chromatogram.

(Peak detection or concentration calculation is not performed.

Chromatograms are not saved.)

Users at Level C or higher are allowed to activate this mode.

Stop

• A mode to stop measurement

• Activate this mode to immediately stop the run mode in the manual

status.

Users at Level C or higher are allowed to activate this mode.