Gc8000-e (iecex), Gc8000-t (tiis) – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 12

Advertising
background image

xi

IM 11B08A01-01E

<GC8000-E

(IECEx)>

IECEx certifi ed: IECEx DEK 11.0083X *

Ex d px II B +H

2

T1 Gb (programmed-temperature oven 320°C max., isothermal oven 225°C

max., liquid-sample valve 250°C max.)
Ex d px II B + H

2

T2 Gb (programmed-temperature oven 225°C max., isothermal oven

225°C max., liquid-sample valve 225°C max.)
Ex d px II B + H

2

T3 Gb (programmed-temperature oven 145°C max., isothermal oven

145°C max., liquid-sample valve 145°C max.)
Ex d px II B + H

2

T4 Gb (programmed-temperature oven 95°C max., isothermal oven 95°C

max., liquid-sample valve 95°C max.)

* :

The symbol “X” used to denote specifi c conditions of use

The symbol “X” is used to provide a means of identifying that essential information for the installation,
use, and maintenance of the equipment is contained within the certifi cate.

- Warning in the label refers you to read the instruction manual for an avoidance of static risk.
- The threaded type and size of the fl ameproof enclosure is only uses “M25x1.5”.
That is the reason why no indication on the fl ameproof enclosure.
On the other hand, the threaded adapter has the indication of its own type and size.
- The gap between fl ameproof joints is different from the standard values of the fl ameproof

standard.

The joint dimensions of the fl ameproof enclosure differ from the minimum or maximum values

asa mentioned in the Ex d standard.

Since the modifi cation is not allowed, this manual gives out no detailed information.
- The instrument modifi cation or parts replacement by any person other than authorized

representative of Yokogawa Electric Corporation is prohibited and will void the fl ameproof
certifi cations.

- All externally powered input signals into the pressurized enclosure shall be isolated by external

relays controlled by the Ex px protection system(Ex px safety device).

<GC8000-T

(TIIS)>

TIIS (The Technology Institution of Industrial Safety)

Ex pd II B +H

2

T1 (programmed-temperature oven 320°C max., isothermal oven 225°C

max., liquid-sample valve 250°C max.)
Ex pd II B +H

2

T2 (programmed-temperature oven 225°C max., isothermal oven 225°C

max., liquid-sample valve 225°C max.)
Ex pd II B +H

2

T3 (programmed-temperature oven 145°C max., isothermal oven 145°C

max., liquid-sample valve 145°C max.)
Ex pd II B +H

2

T4 (programmed-temperature oven 95°C max., isothermal oven 95°C max.,

liquid-sample valve 95°C max.)

2nd Edition : May 11, 2012-00