<2. Installation, Piping, and Wiring>

2-20

IM 11B08A01-01E

Hex.socket set screw

Cable packing adapter

Nut

Connection threads

Case of electronics section
(Pressurized enclosure)

Gasket

Body

Identification mark

Double-deck packing
(Replace it with an appropriate one from the accessory kit.)

Guide ring

Gland

Coupling

Cover

Cable

Hex key wrench

Hex.socket set screw

Plate

This plate is included with shipment.
Please keep the removed the plate.

To the connection port of
the electronics section

Figure 2.14

Procedure for mounting a cable packing adapter for the electronics section

The cable packing adapters comply with the Technical Standard of the Ministry of Health, Labour
and Welfare, Japan.

Table 2.7

Double-deck packing size for electronics section

Connection

port screw

Applicable

cable O.D.

Identifi cation mark

Recommended

torque for the

gland (N·m)

G3/4

ø9.0 to ø10.0

SCFP2010 ø9.0 to ш10.0

25

ш10.0 to ø11.0 SCFP2011 ø10.0 to ш11.0

20

ш11.0 to ø12.0 SCFP2012 ø11.0 to ш12.0

20

ш12.0 to ø13.0 SCFP2013 ø12.0 to ш13.0

20

ш13.0 to ø14.0 SCFP2014 ø13.0 to ш14.0

25

ш14.0 to ø15.0 SCFP2015 ø14.0 to ш15.0

20

ш15.0 to ø16.0 SCFP2016 ø15.0 to ø16.0

20

*:

When a cable O.D. falls under two categories, try both and choose the more suitable one.

For Ethernet cable, use sealing fi ttings in the accessories kit for the connection port (see Figure
2.15). Remove the attached cable packing adapter and mount the sealing fi tting back in place.
Six connection ports are provided in the electronics section. Use convenient ones.

2nd Edition : May 11, 2012-00