<6. Maintenance>

6-11

IM 11B08A01-01E

(7) Remove the screws shown below and remaining connecters, if any, to remove the board

from the assembly.

Screws

(8) Remove the two screws shown below to disconnect the connectors between the boards.

Screws

Safety Board 2

Insulation Sheet

Safety Board 1

(9) After replacing the board, follow steps (5) to (8) in the reverse order.
(10) Confi rm that the connectors are placed in the correct locations.
(11) Install the cover on the protection system. Rotate the cover until it stops, then tighten the

locknut.

<Replacement Parts>

K9800SB

100 V Specifi cation (Safety board 1)

K9800TB

100 V/200 V Specifi cation (Safety board 2)

K9800SD

200 V Specifi cation (Safety board 1)

B. Relays

(1) Stop the operation. (See “3.3.4 Stopping operation” for this procedure).