<Appendix D Action of External Input and Output Signals>

App.D-24

IM 11B08A01-01E

<GC6>
The Analyzer number (e) to be used in the <GC1000/GC8> format is not provided.

(3) Error alarm (fi xed to 45 characters)

This is output only when an alarm occurs.
All the alarms currently occurring are output at the time of startup.

<GC1000/GC8>

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45

A S

/

,

:

,

CR LF

└──┘

└──┘

└──┘

└──┘

└───────┘

└─────┘

a

b

c

d

e

f

a:

Month (01 to 12)

b:

Day (01 to 31)

c:

Hour (01 to 24)

d:

Minute (00 to 59)

e:

Error alarm types

AD1_, AD2_ = Detector 1 (or 2) calibration error

TMPH = Temperature control error or temperature exceeding the higher or lower limit

CAR1, CAR2 = Carrier gas 1 (or 2) pressure drop

EXT1 to EXT8 = External contact inputs 1 to 8

FLM1, FLM2 = FID1 (or 2) extinguished

PRT_ = Calibration repeatability error

CAL_ = Calibration out of range

f:

Analyzer number (1 to 240) (zero suppression provided)

<GC6>
The Analyzer number (f) to be used in the <GC1000/GC8> format is not provided. Error alarm

types (e) are as follows:

e:

Types of error alarms

CAR_ = Carrier gas 1 or 2 pressure drop (Pressure switch 1 or 2 OFF)

FLM_ = FID1 or 2 extinguished

PRT_ = Calibration repeatability error

CAL_ = Calibration out of range

(4) Input Data Format

Stream change command (SM/SE/SC)

When the stream sequence is set for the measurement condition, the command changes

the starting stream number of the current stream sequence. This change can be carried out

in any operation mode (Run, Stop or Pause) except when the operation mode change or the

measurement status change is being requested.
The command changes only the starting stream number, not the stream sequence specifi cation.
The result is written in fl ash memory.

1 2 3 4 5 6 7 8

For GC6

1 2 3 4 5 6

S M

CR LF

S C

CR LF

└──┘

└┘

└──┘

a

b

a

1 2 3 4 5 6 7 8 9 10

S E

CR LF

└──┘

└─────┘

a

b

a:

Starting stream number (1 to 31 or 01 to 31)

b:

Analyzer number (1 to 9 when M comes in the second digit, and 1 to 240 or 001 to 240 when E comes in the second digit)

2nd Edition : May 11, 2012-00