<Appendix C Standard Specifi cations>

App.C-6

IM 11B08A01-01E

FM-Y: Not

required

Note: For installation in hazardous area, an explosionproof

structure authenticated by a relevant certifi ed body
shall be prepared by customer.

The same number unit of the signal interrupter as
contact output card is required.

4.2.6

Contact Input (Option, Up to 4 cards)

Number of contacts: 8 per card, maximum of 32

Function: The following settings are possible at

each contact point.

Alarm:

Occurrence of external contact alarm

(level 2 or 3)

Stream sequence: Performing the specifi ed

stream sequence

Stream (continuous):

Measuring the

specifi ed stream continuously

Stream (once):

Measuring the specifi ed

stream one time

Calibration/Validation: Calibrating

or

validating the specifi ed stream

Operation mode:

Changing the operation

mode (run, pause, stop)

Range change:

Changing the stream and peak

ranges

Contact specifi cations:

Zero voltage

contact input

Contact rating:

5 V DC, 20 mA or more

Input signal:

Open signal: Input load 100 kΩ or more

Close signal:

Input load 200 Ω or less

Operation on input: NC or NO (selectable)

4.2.7

Contact Input/Output (Option, Up to 4 cards)

Number of contacts:

3 for input and 3 for output per card,

maximum of 12 for input and output

each

Function, Contact specifi cation, Contact rating,

Operation specifi cation:

Contact output:

The same functions as in “4.2.5

Contact Output”

Contact input:

The same functions as in

“4.2.6 Contact Input”

External I/O Cutoff Output:

Number of outputs: 1

Function: Monitoring the purge air pressure

in the electronics section, applying

power (24 V DC) to the signal

interrupter when the state is normal.

Signal interrupter (Rack-mounted type, AC:

K9806AN, DC:K9806AJ):

FM-X, ATEX, IECEx, TIIS:

Additionally required. Signal is

interrupted by power supply OFF

signals from the external I/O cutoff

output (contact output only).

FM-Y: Not

required

Note: For installation in hazardous area, an explosionproof

structure authenticated by a relevant certifi ed body
shall be prepared by customer.

The same number unit of the signal interrupter as
contact output card is required.

4.3 Air output

Atmospheric balance valve:

Number of points:

Maximum of 6 (1 for a GCM, up to 2

for an isothermal oven)

Pressure: 350 kPa

Connection:

6 mm or 1/4″ tube, the

pressure control section of the oven

unit

Stream switching valve:

Up to eight automatic valve streams:

Number of points: Maximum of 8 points

Pressure: 350 kPa

Output method:

1 to 1 output

Nine or more automatic valve streams (Note:

Applicable only to 1GCM):

Number of points: 4 points (4 bits, 1 to 15

streams) or 5 points (5 bits, 16 to 31

streams)

Pressure: 350 kPa

Output method:

Binary code output

Connection:

6 mm or 1/4″ tube, the pressure

control section of the oven unit

4.4 User Programming (Option, Up to 1 unit)

The following functions can be set:

• Calculation of analysis results

• Changing operation mode

• ON/OFF operation of DO

• Reading states of DI, AI

5. Communication converter/Signal interrupter

Weight: Approximately

500g

Installation location: Non hazardous area (For

installation in hazardous area, an

explosionproof structure authenticated

by a relevant certifi ed body shall be

prepared by customer.)

Ambient condition during operation:

-10 to 50°C, 95 % RH or less (no

condensation)

Ambient condition during storage:

-40 to 85°C, no condensation

Safety standard: CE Compliance

5.1 Communication converter

• RS-422/RS-232C converter:

(Rack-mounted type: K9806AS, Desktop

type: K9806AT)

Number of port:

1

Communication speed:

maximum of

38400bps

Power supply:

24 V DC (Supplied from

External I/O Cutoff Output of Serial

communication card in GC8000)

Earth: Functional

earth

Other function:

Signal interrupter function

• Ethernet/SHDSL converter: K9802PB

Number of port:

1

Communication speed:

maximum of

5696kbps

2nd Edition : May 11, 2012-00