App.a-4 – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 433

Advertising
background image

<Appendix A Principle of Gas Chromatograph>

App.A-4

IM 11B08A01-01E

FPD

Figure 5 shows the structure of the FPD. As the measured gas containing a sulfur component is

led into the excess hydrogen flame, the component containing the sulfur atoms is excited. The

FPD detects the luminous intensity of the light emitted when this excited component returns to its

base state using a multiplier phototube and converts it to a voltage.
This voltage represents the concentration of the sulfur component in the measured gas.
The FPD can measure the sulfur component with a high sensitivity of 0.2 ppm.

F_A05.ai

Photomultiplier tube

Hydrogen flame

Jet pipe (nozzle)

Carrier gas

+

Sampling gas

Air fo combustion

Hydrogen gas for combustion

Hydrogen gas

for combustion

Figure 5 Basic Confi guration of Flame Photometric Detector

2nd Edition : May 11, 2012-00

Advertising