Detector type: tcd - execution screen – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 285

Advertising
background image

<5. EtherLCD>

5-34

IM 11B08A01-01E

Detector type: TCD - Execution screen

Press F2 (Adjust).

1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5

Det Adjustment

Oven #

1 Det # 1 ( TCD )

>

- Det operation Off

Voltage value 0.0 V
Cuttent value 0.0 mA
- Target current

Main

<<

<

>

>>

F1

F2

F3

F4

F5

F6

Figure 5.45

Example of the detector type: TCD - execution screen

F2 (Main):

Displays the Det Adjustment (Reset) screen.

F3 (<<):

Decreases the value by 16 steps within the range of 0 to 255 (for the TCD, by
approximately 0.94 V within the range of 5.0 V to 20.0 V).

F4 (<):

Decreases the value by 1 step within the range of 0 to 255 (for the TCD, by
approximately 0.06 V within the range of 5.0 V to 20.0 V).

F5 (>):

Increases the value by 1 step within the range of 0 to 255 (for the TCD, by
approximately 0.06 V within the range of 5.0 V to 20.0 V).

F6 (>>):

Increases the value by 16 steps within the range of 0 to 255 (for the TCD, by
approximately 0.94 V within the range of 5.0 V to 20.0 V).

• To toggle the Det On/Off status, set the cursor on the Det operation and press the Set/Ent

key. The screen displays “Setting succeeded” or “Setting failed” in the second line on the
bottom.

• If the Up/Down command keys are pressed to adjust the value within the range of 0 to 255

and the command is successfully sent internally, “Setting succeeded” is displayed. Users
can confi rm the results with the voltage value and current value refreshed every second.

• Selecting the Target current displays Target current (AAA/BB):. If the value 250/10 is

entered, it will be shown as 250+/–10 mA, 250 as 250+/– mA and 50/5 as +/–5 mA.

• The numbers of ovens to which no detectors are installed cannot be set via F5.

TIP

The Det operation, Voltage value and Current value are refreshed at a one-second interval.