<Appendix D Action of External Input and Output Signals>

App.D-1

IM 11B08A01-01E

Appendix D Action of External Input and

Output Signals

Analog output

There are two types of analog output. The maximum number of outputs depends on the type of

outputs.

Analog hold output

Output of analysis results

Up to 32 outputs (8 per card)

The output range can be arbitrarily set within the measurement range.

Note: cannot coexist with GCCU-MarkII (serial) communication.

Chromatogram output Output of chromatograms

Up to 8 outputs

The same detector number can be assigned to multiple A/Os.

The auto gain and auto zero can be set for each detector.

Note: can coexist with GCCU-MarkII (serial) communication.

(1) Analog Hold Output

This is a function to output analysis results as A/O.
The A/O range can be arbitrarily set within the measurement range in the individual peak setting

by using partial settings.

Some partial settings of the measurement range and A/O values are shown in Table 1.

Table 1

Individual peak setting

A/O range

Concentration

(ppm)

A/O value

Stream

- peak

Measurement

range (ppm)

Partial setting Concentration

(ppm)

Current

(mA)

Voltage

(V)

Current

(mA)

Voltage

(V)

#1-#1

1000

1.000

1000.000

20

5

800.000

16.8

4.2

0.000

0.000

4

1

#2-#1

1000

0.800

900.000

20

5

800.000

14.7

3.7

0.500

600.000

4

1

*

Voltage of 1 to 5 V is converted from electric current of 4 to 20 mA using a 250-ohm shunt resistor.

In setting the analog hold output, the number of output analysis values and output actions are

different when an actual stream is set and when “99” is set for the output stream number. Actions

of contact output are different whether a contact output is used or not as a stream identifi cation

signal.

Analog hold output

When an actual stream number is set for

the output stream number, up to 2 analysis

values (different streams) are output.

Without a stream identifi cation signal
With a stream identifi cation signal

When “99” is set for the output stream

number, an analysis value is output for every

stream that belongs to the GCM.

With a stream identifi cation signal

(No other option)

A stream identifi cation signal is a binary-coded stream number which is encoded by turning on

and off multiple contact outputs. To use the stream identifi cation signal as a stream identifi cation

fl ag, which informs upper systems of the update of A/O, as many D/Os as the streams are

required to be initially prepared.
For calibration (or validation) streams, the stream identifi cation signal output is initially set to

“Exist” or “None.” It is set to “Exist” at shipment unless otherwise specifi ed.

• Output exists for calibration (or validation)
• No output for calibration (or validation):

2nd Edition : May 11, 2012-00

App.D