<5. EtherLCD>

5-50

IM 11B08A01-01E

5.4.5 Method

Setup

GCM method setup screen (Main)

1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5

GCM Method Setup

GCM #1 / Method #1
>

- Main cycle 60.0 s

- Warming up time 5.0 s
- Stream valve ON time

-5.0 s

- Stream valve OFF time 0.0 s
- Pause time 55.0 s

Main

Seq

Sys

Copy

GCM# Method#

F1

F2

F3

F4

F5

F6

Figure 5.63

Example of the GCM method setup screen

F1(Menu):

Displays the Table Menu screen.

F2(Seq):

Displays the GCM Sequential screen.

F3(SYS):

Displays the SYS Method screen.

F4(Copy):

Displays the GCM Method Copy screen.

F5(GCM #): Specifi es the GCM number.
F6(Method #): Specifi es the Method number.

• Pressing F3 (SYS) displays the SYS Method Setup screen with the smallest number among

the SYS numbers set to GCM.

• Pressing F5 (GCM #) displays GCM number: in the second line on the bottom. Entering the

method number refl ects the GCM number in the second line on the top and displays the
GCM method setup assigned to that number.

• Pressing F6 (Method #) displays Method number: in the second line on the bottom. Entering

the method number refl ects the method number in the second line on the top and displays
the GCM method setup assigned to that number.

Display Item

Name

Set Lower Limit Upper Limit

Unit

Remarks

GCM #

GCM number

1

6

Method #

Method number

1

6

Main cycle

10

See right

seconds Upper Limit will depend on

sample rate

When 20 ms: 2700 seconds

When 40 ms: 5400 seconds

When 80 ms: 10800 seconds

When 160 ms: 21600 seconds

* The smallest sample rate set

for the detector used by the SYS

belonging to the relevant GCM.

Warming up

time

5.0

9999.9

seconds

Stream valve

ON time

−Warming

up time

0.0

seconds

Stream valve

OFF time

0.0

Main cycle

− 2

seconds

Pause time

5

Main cycle

− 2

seconds

2nd Edition : May 11, 2012-00