<Appendix E Computation Scheme>

App.E-4

IM 11B08A01-01E

(1) Signal Level Processing

There are three kinds of processing: base level setting, level detection, and noise detection.

Base level setting

Level detection

Noise detection

Detector signal data

Base level value

Signal level value

Noise value

F020501.ai

Select any one of these.

Base Level Setting

The signal level in the voltage at the specifi ed time in the analysis cycle (operation time) is stored

as the base level value.

Level Detection

The difference between the latest base level and the signal level at the specifi ed time in the

analysis cycle (operation time) is detected as the signal level.

F02050102.ai

Base level setting
Base level value = E1 mV

E1 mV

E3 mV
0 mV

Peak height from the base level

Level detection
Level value = E2 mV

Base level setting
Base level value = E3 mV

Noise Detection

The noise value of the signal at the specifi ed time in the analysis cycle is obtained.
The difference between the maximum and minimum values among 20 data is converted into a

value per second as the noise value.

F02050103.ai

n=20

Maximum and
minimum value