App.d-17, App.d, Stream – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 462: Range switching

Advertising
background image

<Appendix D Action of External Input and Output Signals>

App.D-17

IM 11B08A01-01E

To give a command to all GCMs, set “0” for the GCM number.

1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5

D/I Setup

Slot #

1 DI # 1

>

- Signal name G1-Seq1

- Process assign Stream sequence
- GCM number 1
- Stream sequence 1

Menu

Status

Slot#

DI#

F1

F2

F3

F4

F5

F6

Figure 22 Process Specifi cation: Stream Sequence (D/I Setting Screen on GC-HMI EtherLCD)

Stream

The Stream function of D/I is a function to give a command to change the measurement status so

that the specifi ed stream is measured a preset number of times using contact input as a trigger.

To measure the specifi ed stream continuously, set “0” for the number of times of measurement.

1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5

D/I Setup

Slot #

1 DI # 1

>

- Signal name STR1

- Process assign Stream (cont)
- Stream number 1 (GCM 1)
- Number of times 0

Menu

Status

Slot#

DI#

F1

F2

F3

F4

F5

F6

Figure 23 Process Specifi cation: Stream (D/I Setting Screen on GC-HMI EtherLCD)

Range Switching

The Range switching function of D/I is a function to give a command to change the range to a

specifi ed range for the range-specifi ed peak of the specifi ed stream using contact input as a

trigger.
To enable the Range switching function of D/I, individual peak setting and multi-range settings

are required in advance for the relevant stream and peak.