<1. Overview>

1-51

IM 11B08A01-01E

Run

W1

S1

W1

S1

W1

[2E2]

Time

Stream specification 1 (Continuous)

Stop command

Stop

● In the case where a stop command is received during the run mode in the stream

specifi cation status (for N times of measurement) [3E2]

The operation mode is immediately prepared to be changed to the stop mode.
The operation mode changes to the stop mode in the stream specifi cation status (for

N times of measurement) when the measurement of the currently-measured stream is
completed.

In this case, the number of times of measurement is counted again from the beginning.
The stop command can be canceled with a cancel command until the operation mode

changes to the stop mode.

Run

W1

S1(2/8)

W1

S1(1/8)

W1

S1(2/8)

W1

S1(1/8)

W1

[3E2]

Time

Stream specification 1 (8 times)

Stop command

* The number of times of measurement is reset.

Run command

Run command

Stop

Run

● In the case where a stop command is received during the pause mode in the stream

specifi cation status [2E6]

The operation mode is immediately changed to the stop mode in the stream specifi cation

status.

In the stream specifi cation status (for N times of measurement), the number of times of

measurement is counted again from the beginning.

Pause

[2E6]

Time

Stream specification 1

Stop command

Stop

● In the case where a forced stop command is received during the run mode in the stream

specifi cation status [2F2]

The operation mode is immediately changed to the stop mode in the stream specifi cation

status.

In the stream specifi cation status (for N times of measurement), the number of times of

measurement is counted again from the beginning.

2nd Edition : May 11, 2012-00

1