<4. GC-HMI (touch panel)>

4-34

IM 11B08A01-01E

Ref stream/Peak set

Set the reference stream number and reference peak number for the peak of the specifi ed
stream number, SYS number, and peak number.

Figure 4.42

Specifi c items to be set (peak 1 - ref stream number/peak 2 - ref stream number /peak 1 - ref
peak number/peak 2 - ref peak number) depend on the process assignment defi ned by stream
number/peak number.

Process assignment is external cubic: Peak 1 only
Process assignment is external linear: Peaks 1 and 2
Otherwise: No

peak

Setting range:

Peak 1 - Ref. stream number: 1 to 31
Peak 2 - Ref. stream number: 1 to 31
Peak 1 - Ref. peak number: 1 to 999 (only the peak number assigned to the reference stream

number can be selected)

Peak 2 - Ref. peak number: 1 to 999 (only the peak number assigned to the reference stream

number can be selected)

Gate /Integ. Set

Set the gate cut method and integration method for the peak of the specifi ed stream number,
SYS number, and peak number.

2nd Edition : May 11, 2012-00