App.e-13, 3) deviation component computation, App.e – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 486: F) total compensation, G) variation coeffi cient computation

Advertising
background image

<Appendix E Computation Scheme>

App.E-13

IM 11B08A01-01E

The computation formula is:

c

i

=

(f

j

s

j

)

f

i

s

i

k

j=l

Where:

c

i

: concentration of peak i

s

i

: area of peak i

f

i

: sensitivity correction coefficient of peak i

(f) Total

Compensation

Total compensation computes the concentration measured in a stream so that the total value of

the measurement components is the total value specifi ed in each stream.
The computation formula is:

c

ni

= c

total

c

j

c

i

k

j=l

Where:

c

ni

: result of total compensation for peak i (excluding the concentrations

of the “Total compensation disable” peaks)

c

ni

: concentration of peak i

c

total

: total of the concentration values

(g) Variation Coeffi cient Computation

Variation coeffi cient computation computes the variation coeffi cient for concentration.
The computation formula is:

c

avrg

=

n

c

i

c

v

=

1

c

span

(c

i

–c

avrg

)

2

n

Where:

c

i

: concentration of peak i

c

span

: measurement span of peak i

c

v

: variation coefficient of concentration for peak i

(3) Deviation Component Computation

Deviation component computation computes the difference between the concentration values

measured in a stream and the total concentration specifi ed for each stream.
Deviation coeffi cient computation can be made for obtained concentration values. See the

section on variation coeffi cient computation in peak processing.
If the result of a computation is negative, the result is counted as 0.

2nd Edition : May 11, 2012-00

App.E

Advertising