<5. EtherLCD>

5-33

IM 11B08A01-01E

5.3.9 Detector

Adjustment

Detector type: TCD - Reset screen

1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5

Det Adjustment

Oven #

1 Det # 1 ( TCD )

>

- Det operation Off

- Reset status Initialized

Menu

Adjust

Reset

Oven#

Det#

F1

F2

F3

F4

F5

F6

Figure 5.44

Example of the detector type: TCD - reset screen

F1 (Menu):

Displays the Table Menu screen.

F2 (Adjust):

Displays the Det Adjustment (Execution) screen.

F4 (Reset):

Initializes voltage values.

Pressing F4 displays the options Abort and Initialize. If the Set/Ent key is
pressed after the selection of Initialize, the TCD voltage value is set to its
minimum value 5.0 V.

F5 (Oven #): Sets an oven number.
F6 (Det #):

Sets a detector number.

• The TCD voltage value is read at a one-second interval, when, if the value is 0.0 V, the

status is refreshed with Initialized; otherwise Setting done.

• The Det operation is refreshed at a one-second interval.
• The numbers of ovens to which no detectors are installed cannot be set via F5.

Display Item

Name

Set Lower Limit Upper Limit

Unit

Remarks

Oven #

Oven number

1

3

Det #

Detector number

1

2

Detector type

TCD

Det operation

0: Off

1: On

* For this operation, the user

level C or higher is required and

the operation mode must be

Stop in Manual status.

Reset status

0: Initialized

1: Setting done

* If the voltage value is

zero, Initialized is displayed;

otherwise Setting done.

2nd Edition : May 11, 2012-00

5