<1. Overview>

1-65

IM 11B08A01-01E

Run

W7

W7

W8

W1

S1

S1

S8

W1

W1

S1

S7

S7

Time

Stream specification 1 (Continuous) Calibration 1

Stream specification 1 (Continuous)

Assinning calibration 1

● In the case where a (semiautomatic) calibration/validation command is received during the

run mode in the stream specifi cation status (for N times of measurement)

Calibration (or validation) is started when the Nth round of the measurement of the currently-

measured stream is completed.

Users at Level C or higher can cancel the calibration/validation command until the

measurement of the currently-measured stream is completed.

When calibration (or validation) is completed for all the streams specifi ed for the selected

calibration (or validation) number, the operation mode changes to the stop mode in the
stream specifi cation status (for N times of measurement).

Run

W7

W7

W8

S8

W1

S1

W1

S1

S7

S7

Time

Stream specification 1 (Twice)

Calibration 1

Stream specification 1 (Twice)

Assinning calibration 1

Run command

Stop

Stop

● In the case where a (semiautomatic) calibration/validation command is received during the

stop mode in the stream specifi cation status

Calibration (or validation) is immediately started.
When calibration (or validation) is completed for all the streams specifi ed for the selected

calibration (or validation) number, the operation mode returns to the stop mode in the
stream specifi cation status.

Run

W7

W7

W8

S8

S7

S7

Time

Stream specification 1

Calibration 1

Stream specification 1

Assinning calibration 1

Stop

Stop

● In the case where a (semiautomatic) calibration/validation command is received during the

pause mode in the stream specifi cation status

Calibration (or validation) is immediately started.
When calibration (or validation) is completed for all the streams specifi ed for the selected

calibration (or validation) number, the operation mode changes to the stop mode in the
stream sequence status.

Run

W7

W7

W8

S8

S7

S7

Time

Stream specification 1

Calibration 1

Stream specification 1

Assinning calibration 1

Pause

Stop

2nd Edition : May 11, 2012-00