<5. EtherLCD>

5-88

IM 11B08A01-01E

Display Item

Name

Set Lower Limit Upper Limit

Unit

Remarks

Slot #

Slot number

1

5

AI #

A/I input number

1

4

Signal Name A/I signal name

Alphanumeric: 8 characters

Filtering const A/I fi ltering constant

0.001

1.000

5.4.19 Network

Setup

Set up IP addresses and networks.

1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5

Network Setup

Main
>

- HY-A IP address 192. 168. 1. 1

- Gateway IP_A 192. 168. 1. 250
- Subnet mask

255. 255. 255. 0

- PCAS confi g Single
- ASGW confi g Single
- DCS confi g Single
- Scaling coef Real

Menu

Redundant

Dial_ip

F1

F2

F3

F4

F5

F6

Figure 5.100

Example of network setup (Main) screen

F1 (Menu):

Displays the Table menu screen.

F4 (Redundant):

Displays the Redundant setup screen.

• F4 (Redundant) is valid only when the Ethernet (ch2) is installed.

Redundant setup screen

Execute the redundant setup.
Press F4 (Redundant).

1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5

Network Setup

Redundant
>

- Virtual IP address 192. 168. 1. 1

- PHY-B IP address 0. 0. 0. 0
- Gateway IP_B 0. 0. 0. 0
- Diag Interval 500 ms

Menu

Main

Dial_ip

F1

F2

F3

F4

F5

F6

Figure 5.101

Example of network setup (Redundant) screen

F1 (Menu):

Displays the Table menu screen.

F3 (Main):

Displays the Main screen.