2 a/o status – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 257

Advertising
background image

<5. EtherLCD>

5-6

IM 11B08A01-01E

5.2.2 A/O

Status

To display the Analog Output (A/O) status, move the cursor (>) with

or

to the A/O Status

row and press

.

1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5

A/O Status

Slot # 1
AO# Output item

Value

1 Det 1-1 AO Chromatogram

0.000

2 - - - - - - - - - - - - - - - - - - - - - - - - - - 0.000
3 - - - - - - - - - - - - - - - - - - - - - - - - - - 0.000
4 - - - - - - - - - - - - - - - - - - - - - - - - - - 0.000
5 - - - - - - - - - - - - - - - - - - - - - - - - - - 0.000
6 - - - - - - - - - - - - - - - - - - - - - - - - - - 0.000
7 - - - - - - - - - - - - - - - - - - - - - - - - - - 0.000
8 - - - - - - - - - - - - - - - - - - - - - - - - - - 0.000

Menu

Table

Slot#

F1

F2

F3

F4

F5

F6

Figure 5.6

Example of the A/O status screen

F1 (Menu):

Displays the Status Menu screen.

F4 (Table):

Displays the A/O Setup screen.

F6 (Slot #):

Sets a slot number (applicable only when multiple analog output cards are
inserted).

TIP

• The Value data are refreshed automatically at a fi xed interval (1 second).
• On the initial screen, the smallest slot number among those of the A/O cards inserted is displayed.
• For the slot number setting, only slot numbers of A/O cards can be accepted.

If no A/O cards are inserted, the Status Menu screen displays “Not load” on the bottom.

Display Item

Name

Set Lower Limit Upper Limit

Unit

Remarks

Slot #

Slot number

1

5

Absolute slot number

AO #

A/O channel number

1

8

Relative channel number

Stream number on

the output

1

31

Peak number on the

output

1

999

Relative peak number

Output item

A/O output peak

name

8 single-byte alphanumeric

characters

Analysis results on

the output

0.000

0.000

999.999

9999.999

When the unit is %

For other units

* The number of effective digits

is 6.

Analysis value unit

on the output

See the Measuring unit of the

Peak Setup-Specifi c screen.

* Only the upper 5 characters

are displayed.

Detector number

None, 1-1, 1-2, 2-1, 2-2, 3-1, 3-2

* Oven number and relative

detector number

Value

A/O value

-0.25

1.25

If the value stored in the

database is 10000, it is shown

as 1.0000.

2nd Edition : May 11, 2012-00