6 stream setup – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 312

Advertising
background image

<5. EtherLCD>

5-61

IM 11B08A01-01E

5.4.6 Stream

Setup

1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5

Stream Setup

Stream # 1
>

- Name of stream

- Str valve number 0
- Type of stream Measurement
- Normalization * * * * * * * * * *
- Total value 100.0 %
- Base line corr Unexecuted
- GCM number 1
- Method number 1
- SIMDIS calib stream

Menu

Strm#

F1

F2

F3

F4

F5

F6

Figure 5.75

Example of the stream setup screen

F1 (Menu):

Displays Table Menu screen.

F6 (Strm #):

Specifi es a stream number (1 to 31).

• The SimDis calib stream item is displayed only when distillation is set to Yes in the GCM

setup (2/2) screen for the GCM specifi ed by the GCM number. Displays a stream type when
it is a sample run.

• In the case of the area correction percentage for GCM, normalization for the stream

belonging to that GCM is displayed by an asterisk.

Display Item

Name

Set Lower Limit Upper Limit

Unit

Remarks

Stream #

Stream number

1

31

Stream name

Alphanumeric: 8 characters

Stream valve

number

0

31*

0: No stream valve

* Only stream valve number

belonging to the GCM can be

set (GCM setup screen).

Stream type

Measurement, Cal, Val,

Calibration run (only when

distilling), Sample run (only

when distilling)

Normalization

Not executed, Executed

* When the GCM of the

relevant stream is confi gured

with multiple SYS, Executed

cannot be set.

Total value

0.1

100.0

Base line corr

Not executed, Executed

GCM number

1

6

Method

number

1

6

Distillation

confi guration

stream

0

31

2nd Edition : May 11, 2012-00