7) liquid-sample valve (lsv), Seal unit – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 393

Advertising
background image

<6. Maintenance>

6-40

IM 11B08A01-01E

F0233.ai

Rotor section

Rotor section

Rotor section

Bearing

Pin

Pin

Top mark

(b) Seat for switching valve

(a) Seat for sample valve

<For gas sample>

<For liquid sample>

Top mark

Valve seat

Valve seat
for liquid

Valve seat
for gas

Groove for
valve seat clocking

Groove for gland
assembly clocking

Figure 6.33

Valve Seat

(7) Liquid-sample Valve (LSV)

The liquid-sample valve is used when the sample is a liquid. See Figure 6.34. The specifi cations
for the liquid sample valve used are listed in the parts list in the operation data.
Check against the code chart (Table 6.3) and replace as required.

Table 6.3

Model

Suffi x code

Option code

Description

GCLV

•••••••••••••••••••••••••••••••••••••• ••••••••••••••••• Liquid sampling valve

Sample volume -1

-2

-3

-4

-5

-6

•••••••••••••••••

•••••••••••••••••

•••••••••••••••••

•••••••••••••••••

•••••••••••••••••

•••••••••••••••••

0.1 μl

0.25 μl

0.5 μl

1 μl

2 μl

3 μl

Wetted material

S

H

•••••••••••••••••

•••••••••••••••••

SUS316

Hastelloy C

Insert material

M

G

C

J

•••••••••••••••••

•••••••••••••••••

•••••••••••••••••

•••••••••••••••••

SUS316

Glass (straight)

Glass (cup)

Glass (with neck)

Sealing material

R

T

•••••••••••••••••

•••••••••••••••••

Rulon

Tefl on

Column connection

A

B

C

D

•••••••••••••••••

•••••••••••••••••

•••••••••••••••••

•••••••••••••••••

1/8"

1/16"

Capillary

Mega-bore

Split/Leakage back fl ash

N

1

2

•••••••••••••••••

•••••••••••••••••

•••••••••••••••••

None

1/8"

1/16"

Seal unit

(1) Stop the operation. (See “3.3.4 Stopping operation” for this procedure).
(2) Turn off the supply of the sample.
(3) If the sample is liquid, purge the sample line with the purge gas (nitrogen gas or instrument

air).

2nd Edition : May 11, 2012-00