3) fpd – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 427

Advertising
background image

<7. Troubleshooting>

7-14

IM 11B08A01-01E

(3) FPD

Troubleshooting diagram for malfunctioning chromatogram is shown as followings.

F0306.ai

Malfunctioning chromatogram

Flow rate

of combustion gas is as

specified

Yes

No

Yes

No

Yes

No

Yes

No

Yes

No

Yes

No

Yes

No

Yes

Yes

No

TUrn on detector

DET FLAME OUT

is occured

Status of

detector is ON

Adjust flow rate of

combustion gas

Flow rate

of combustion gas is as

specified

Yes

No

Level

of flame signal increase

Adjust flow rate of

combustion gas

Increase combustion air pressure

to twice value as specified

Decrease combustion air

pressure to as specified value

Re-turn on detector with

correct procedure

Voltage value

for ignition at DET board is

as specified

Repalce detector board

Yes

No

Voltage value

for ignition at DET is as

specified

Replace wiring for detector

Replace detector

Baseline of

detector is malfunctioning

Column system failure or

sampling system failure

Temperature of oven is

stable

Wait for stabilization of

oven temperature

Yes

No

Combustion gas is clean

Use clean combustion air

Yes

No

External piping is clean

Use clean piping

No

Replace high voltage

power supply

High voltage

value for amp. board is as

specified

Voltage value

for amp. board at DET board is

as specified

Replace detector board

Yes

No

Voltage value

for amp. board at amp. board is

as specified

Replace wiring for

amp. board

Replace detector card, amp.

board or photo-multiplier

Figure 7.6

Troubleshooting Diagram for Malfunctioning Chromatogram (FPD)