Warning, Atex, iecex – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 133

Advertising
background image

<2. Installation, Piping, and Wiring>

2-16

IM 11B08A01-01E

3/4NPT(F)

9/16 inch

A

3/4NPT

Connection port of the protection system *3

Connection port of the electronics section *1

Electronic section
(Pressurized enclosure)

*1: Any of the six connection ports of the

electronics section can be equally used.

*2: A plug is attached to connector.

Remove and keep it for future use.

*3: In the FM-Y, the protection system is
not

provided.

Protection system *3

(Flameproof enclosure)

as close as possible
not more than 18 inches (457 mm)

18 inches (457mm)
maximum

Sealing Fitting

Conduit

HAZARDOUS
LOCATIONS

NON-HAZARDOUS
LOCATIONS

Detail of A

Connector

Nut

Gasket

Plate

Plug *2

NONHAZARDOUS

LOCATION

EQUIPMENT

NONHAZARDOUS

LOCATION

EQUIPMENT

Figure 2.17

Wiring in FM specifi cation

ATEX, IECEx

WARNING

• All wiring shall comply with Local Electric Codes and Requirements.
• In a hazardous area, use appropriate fl ameproof-certifi ed parts for connecting cables.
• All externally powered input signals into the pressurized enclosure shall be isolated by

external relays controlled by the Ex px protection system(Ex px safety device).

2nd Edition : May 11, 2012-00