Gcm sequential display screen, Stream valve set screen – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 303

Advertising
background image

<5. EtherLCD>

5-52

IM 11B08A01-01E

GCM Sequential Display Screen

Press F2( Seq).

1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5

GCM Sequential Display

GCM #1 / Method #1
Time Action
-5.0 StrV on time
0.0

StrV off time

50.0 S1 Peak detect end
55.0 Pause time

Main

StrV

F1

F2

F3

F4

F5

F6

Figure 5.65

Example of the GCM sequential display screen

F1(Main):

Displays GCM Method (Main) screen.

F3(StrV):

Displays Stream Valve Set screen.

Displays the sequence of the GCM method and the SYS method set in the GCM.

When newly displayed, the sequential display from the fi rst sequence is displayed.
However, SYS sequence set to No for the SYS usage set is not displayed.

Display Item

Name

Set Lower Limit Upper Limit

Unit

Remarks

GCM #

GCM number

1

6

Method #

Method number

1

6

Time
Action Time

Pause time

S1 to S6 Peak detect end

StrV on time, StrV off time

Stream Valve Set Screen

Press F3 (StrV).

1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5

Stream valve Set

GCM #1 / Method #1
>

- Stream valve ON time

-5.0 s

- Stream valve OFF time

0.0 s

Seq

F1

F2

F3

F4

F5

F6

Figure 5.66

Example of the stream valve set screen

F1(Seq):

Displays Sequential Display screen.

2nd Edition : May 11, 2012-00