<Appendix E Computation Scheme>

App.E-5

IM 11B08A01-01E

(2) Peak Processing

Peaks are detected from the detector signal data, and the concentration is calculated from the

area or height of the peak.

Detector signal data

Time gating

Retention-time

detection

Tangent

processing

Square

processing

F020502_1.ai

Peak detection processing

Select whether to execute

Auto-tracking

Select whether to execute

Select one of the following

Select whether to execute

Peak-on time/peak-on level

Retention time

Peak height

Peak-off time/peak-off level

Result of tangent correction

Tailing coefficient

Peak half width

Peak

waveform

Area

integration

Area calculation

by tangent

skimming

Peak waveform processing

Peak area

Slope gating

Zone gating

Slope/time

gating

Time/slope

gating

Base line

correction

Area calculation

by vertical

separation

2nd Edition : May 11, 2012-00