Detector type: fid-mc - reset screen – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 286

Advertising
background image

<5. EtherLCD>

5-35

IM 11B08A01-01E

Display Item

Name

Set Lower Limit Upper Limit

Unit

Remarks

Oven #

Oven number

1

3

Det #

Detector number

1

2

Detector type

TCD

Det operation

Off, On

* For this operation, the user

level C or higher is required and

the operation mode must be

Stop in Manual status.

Voltage value

0.0

20.0

V

Current value

0.0

500.0

mA

Target current

Numeric, period “.” and slash “/”

are enabled: 16 characters

Input format: AA.A/B.B

Output format: AA.A+/–B.B mA

Detector type: FID-MC - Reset screen

Press F2 (Adjust).

1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5

Det Adjustment

Oven #

1 Det # 1 ( FID-MC )

>

- Det operation Off

- Reset status Initialized

Menu

Adjust

Reset

Oven#

Det#

F1

F2

F3

F4

F5

F6

Figure 5.46

Example of the detector type: FID-MC - reset screen

F1 (Menu):

Displays the Table Menu screen.

F2 (Adjust):

Displays the Det Adjustment (Execution) screen.

F4 (Reset):

Initializes a voltage value.

Pressing F4 displays the options “Abort” and “Initialize.” If the Set/Ent key
is pressed after the selection of “Initialize,” the MC voltage value is set to its
minimum value 10.0 V.

F5 (Oven #): Sets an oven number.
F6 (Det #):

Sets a detector number.

• Pressing F4 displays the options “Abort” and “Initialize” in the second line on the bottom. If

the Set/Ent key is pressed after the selection of “Initialize,” the MC voltage value is set to its
minimum value 10.0 V.

• The MC voltage value is read at a one-second interval, when, if the value is 0x0, the status

is refreshed with “Initialized”; otherwise Setting done.

• The Det operation is refreshed at a one-second interval.
• The numbers of ovens to which no detectors are installed cannot be set via F5.

2nd Edition : May 11, 2012-00