10 cal/val setup – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 324

Advertising
background image

<5. EtherLCD>

5-73

IM 11B08A01-01E

Display Item

Name

Set Lower Limit Upper Limit

Unit

Remarks

Stream #

Stream number

1

31

Use num #

Use

0

999

Number of peaks in the relevant

stream

Assign #

Assign

0

999

Number of assigned peaks in

the relevant stream

Peak Relevant

stream

peak number

1

999

Name

Peak/Processing

name

Eight one-bite characters

Usage

Peak usage status

ON when in use, OFF when not

in use

Process

(Type)

5.4.10 Cal/Val

Setup

1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5

Cal/Val Setup (Main)

GCM # 1
> - Cal(Val) method Semi-auto
- Auto start date

2011/11/22

- Auto start time 00:00
- Time interval 000 day:00:05
Auto cal status Stopping

Menu

Start

Stop

Cal

Val

GCM#

F1

F2

F3

F4

F5

F6

Figure 5.81

Example of Cal/Val Set (main) screen

F1 (Menu):

Displays the Table menu screen.

F2 (Start):

Executes the Auto cal start command.

F3 (Stop):

Executes the Auto cal stop command.

F4 (Cal):

Displays the Calibration set screen.

F5 (Val):

Displays the Validation Set screen.

F6 (GCM #): Specifi es a GCM Number.

• In the case of auto cal, when auto calibration is started using the F2 (Start) auto

calibration start command, the message “Operation change selected” is displayed. When
measurement status is executing calibration, validation or auto-calibration status, the
message “Operation change not selected” is displayed.

• In the case of auto cal, auto cal is stopped using the F3 (Stop) auto cal stop command, the

message “Operation change selected” is displayed. When measurement status is stopping
calibration, validation or auto-calibration status, the message “Operation change not
selected” is displayed.

• When auto-calibration start command is executed using F2 (Start), and where Cal/Val

setting for auto cal and auto val is not set to execute, the message “Auto cal or val has not
been set yet” is displayed.

• When auto cal status is executing and Cal (Val) method is changed to semi-auto or manual,

the auto cal status is changed to stopping. However, for calibration that is already in session,
the analysis is performed until the main cycle ends.

2nd Edition : May 11, 2012-00