Instrument modifi cation or part replacement must be conducted by the personnel of Yokogawa or
its assigned representative only. The contents of this chapter must be conducted by Yokogawa
service staff or a person with a proper training.

6.1

Maintenance and Inspection

6.1.1 Explosion-protection

Structure

When performing maintenance, fi rst inspect the system according to the instructions given in
Table 6.1 before operating, in order for the explosionproof performance to be maintained.
If the following damages occurred, contact Yokogawa service center:
(1) Damage around the screws on the protection system (fl ameproof structure).
(2) Damage to any enclosure surface or transparent part.
(3) Any crack or excessive deformation to any of the gaskets.

Table 6.1

Maintenance and Inspection

Item

Method

Criterion

Corrective Action

Enclosure

Visual

No damage

Contact YOKOGAWA

Touch panel

surface

Visual

No damage

Contact YOKOGAWA

Setscrew

Visual and by touch No looseness, dust or rust

Retighten and clean, and

contact YOKOGAWA

Gaskets

Visual

No cracks or excessive deformation

Contact YOKOGAWA

Cable inlet

Visual and by touch No damage, deterioration or looseness Retighten or replace

Terminal

Visual and by touch No looseness or dirt on insulators

Retighten , parts change and

clean

6.1.2

Stopping the system

Unless in case of an emergency, stop the system according to “3.3.4 Stopping operation” and
then turn off the power.

6.1.3 Routine

Inspection

The routine maintenance inspection must be done according to Table 6.2. In the inspection,
confi rm that all of the readings are consistent with the values indicated in the operation data.
These readings, if previously entered on the Operation Condition display menu, can also be
confi rmed on the screen as well.
Since these inspection items vary depending on the instrument specifi cations, select as required
according to the specifi cations.