Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 90

Advertising
background image

<1. Overview>

1-63

IM 11B08A01-01E

11/03/22

15:15:45

Cal/Val Set (Main)

GCM #1
- Cal(Val) method

Semi-auto

- Auto start date

2011/07/27

- Auto start time

09:00

- Time interval

001 day:00:00

Auto cal status

Stopping





Menu

Start

Stop

Cal

Val

GCM#

F1

F2

F3

F4

F5

F6

Figure 1.29

An example of the calibration/validation setting screen on the GC-HMI EtherLCD.

F0125.ai

Figure 1.30

An operation example of the calibration (validation) window on the GCM operation

status screen on the GC-HMI analyzer operation display

● In the case where a (semiautomatic) calibration/validation command is received during the

run mode in the stream sequence status

Calibration (or validation) is started when the measurement of the currently-measured

stream is completed.

2nd Edition : May 11, 2012-00

1

Advertising