Simdis calib data setup – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 350

Advertising
background image

<5. EtherLCD>

5-99

IM 11B08A01-01E

Display Item

Name

Set Lower Limit Upper Limit

Unit

Remarks

GCM #

GCM Number

Temp Unit

0) °C

1) Fahrenheit (Refer to SI unit

limitation)

Det number of

SIMDISM

0) NA

1) 1-1

2) 1-2

3) 2-1

4) 2-2

5) 3-1

6) 3-2

Only detectors assigned to

relevant GCM can be set up.

Response factor

updating

Not done, Done

Section updating

Not done, Done

Calilb curve

updating

Not done, Done

SimDis Calib Data Setup

Press F4 (Curve).

1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5

SimDis Calib Data Setup

(1/3)

1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5

SimDis Calib Data Setup

(2/3)

GCM #

1 / SYS # 1 / Method # 1

GCM #

1 / SYS # 1 / Method # 1

> - Data acquisition str 0

>

- Data acquisition peak3 0

- Calib curve coef a 1.000

- Elapsed time 3 0.0 s

- Calib curve coef b 1.000

- Std B.P 3 0.0

- Data acquisition peak1 0

- Data acquisition peak4 0

- Elapsed time 1 0.0 s

- Elapsed time 4 0.0 s

- Std B.P 1 0.0

- Std B.P 4 0.0

- Data acquisition peak2 0

- Data acquisition peak5 0

- Elapsed time 2 0.0 s

- Elapsed time 5 0.0 s

- Std B.P 2 0.0

- Std B.P 5 0.0

Menu

SYS# Method#

Menu

SYS# Method#

F1

F2

F3

F4

F5

F6

F1

F2

F3

F4

F5

F6

Figure 5.112

Example of SimDis Calib Data Set (1/3) screen

F1 (Dist):

Displays the SimDis Setup screen.

F5 (SYS #):

Specifi es a SYS number.

When F5 (SYS #) is pressed, “SYS No.:” is displayed on the second line below. Use the numeric
keypad to enter the number and press the SET/ENT key, and a cursor jumps to the SYS number.
Only the SYS numbers set in GCM is selectable. When any other number is entered, “Over limit”
is displayed on the second line below.

• F6 (Method #): Specifi es a Method number.

• The GCM number displayed on the second line above is the same as the one set on the

SimDis Setup screen.

• Data acquisition streamSet the cal run stream number acquired for acquiring data to

compile a SimDis calib curve. On the stream setup screen, when a stream number with
other stream type than Cal run is about to be set, the error message “No cal run stream” is
displayed on the second line below.

• Calib curve coef a & b
Linear ramp a & slice b of a SimDis Calib curve. Auto updating and manual entry are

possible.

2nd Edition : May 11, 2012-00

5