Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 72

Advertising
background image

<1. Overview>

1-45

IM 11B08A01-01E

● In the case where a forced stop command is received during the pause mode in the stream

sequence status [1F6]

The operation mode is immediately changed to the stop mode in the stream sequence

status.

Pause

[1F6]

Time

Stream sequence 1

Forced stop command

Stop

● In the case where a pause command is received during the stop mode in the stream

sequence status [1H1]

The operation mode is immediately changed to the pause mode in the stream sequence

status.

Pause

Stop

Run

W2

W3

S2

S1

W1

W2

S1

S1

[1H1]

Time

Stream sequence 1

Pause command

Run command

Pause time

● In the case where a pause command is received during the run mode in the stream

sequence status [1H2]

The operation mode is immediately prepared to be changed to the pause mode in the

stream sequence status.

The operation mode changes to the pause mode in the stream sequence status at a preset

time during the measurement of the currently-measured stream.

The pause command can be canceled with a cancel command until the operation mode

changes to the pause mode.

● In the case where a run command is received during the pause mode in the stream

sequence status [1G6]

The suspended measurement is resumed.

Run

Run

Pause

Rest of W2

Rest of S1

W1

W2

W3

S2

S3

S1

W3

S2

W1

W2

S1

[1G6]

Time

Stream sequence 1

Run command

[1H2]

Pause command

2nd Edition : May 11, 2012-00

1