<1. Overview>

1-67

IM 11B08A01-01E

When the measurement status is set to the calibration (or validation) status, commands to

start/stop automatic calibration are invalid.

Users at Level C or higher are allowed to specify the setting and issue commands.
• When all the calibration (or validation) patterns of which automatic execution is enabled

are completed, the measurement status returns to the status before the calibration (or
validation). The operation mode after calibration (or validation) is determined as follows
depending on the measurement status and the operation mode before the calibration (or
validation).

Measurement status and operation mode before calibration (or

validation)

Operation mode after

calibration (or validation)

Stream sequence/Stream specifi cation (for continuous measurement), Run

Run

Stream specifi cation (for N times of measurement), Run

Stop

Stream sequence/Stream specifi cation (for continuous measurement)/

Stream specifi cation (for N times of measurement), Stop

Stop

Stream sequence/Stream specifi cation (for continuous measurement)/

Stream specifi cation (for N times of measurement), Pause

Stop

F0131.ai

Figure 1.31

An example of the calibration (validation) method window on the overall GCM

information screen on the GC-HMI analyzer operation display

2nd Edition : May 11, 2012-00