Installation site and environment

The analyzer specifi cations allow it to be used in hazardous areas as defi ned by DIV1, GPS B, C
& D, T1, T2, T3, T4 (FM) or Zone 1 IIB + H2T1, T2, T3, T4 (ATEX, IECEx, TIIS). However, never
install the analyzer in an area where dense explosive gas exists for a long time.
For the class of hazardous areas:
For FM, refer to Article 500 of the National Electrical Code (NEC).
For ATEX or IECEx, see IEC 60079-10-1.
For TIIS explosion protection, refer to Article 1 (15) to (17) of the Constructional Requirements for
Electrical Equipment for Explosive Atmospheres (Japanese only).

Wiring works

<FM>

WARNING

• All wiring shall comply with National Electric Code ANSI/NFP A 70 and Local Electric Codes.
• In a hazardous area, use conduits for wiring in the protection system or to electronics

sections.

CAUTION

• The unused electrical connection ports should be closed with an appropriate fl ameproof-

certifi ed plug.

• Analyzers have pressurized enclosures. The cable end should be sealed in order to

maintain pressure to the pressurized enclosure. Otherwise, power is not supplied to the
electronics section.

<ATEX, IECEx>

WARNING

• All wiring shall comply with Local Electric Codes and Requirements.
• In a hazardous area, use appropriate fl ameproof-certifi ed parts for connecting cables.
• All externally powered input signals into the pressurized enclosure shall be isolated by

external relays controlled by the Ex px protection system(Ex px safety device).

CAUTION

• The unused electrical connection ports should be closed with an appropriate fl ameproof-

certifi ed plug.

• Analyzers have pressurized enclosures. The cable end should be sealed in order to

maintain pressure to the pressurized enclosure. Otherwise, power is not supplied to the
electronics section.