Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 91

Advertising
background image

<1. Overview>

1-64

IM 11B08A01-01E

Users at Level C or higher can cancel the calibration/validation command until the

measurement of the currently-measured stream is completed.

When calibration (or validation) is completed for all the streams specifi ed for the selected

calibration (or validation) number, measurement is resumed from the next stream in the run
mode in the stream sequence status.

Run

W7

W7

W8

W3

S2

S3

S8

W1

W2

S1

S7

S7

Time

Stream sequence 1

Calibration 1

Stream sequence 1

Assinning calibration 1

● In the case where a (semiautomatic) calibration/validation command is received during the

stop mode in the stream sequence status

Calibration (or validation) is immediately started.
When calibration (or validation) is completed for all the streams specifi ed for the selected

calibration (or validation) number, the operation mode returns to the stop mode in the
stream sequence status.

Stop

Stop

Run

W7

W7

W8

S8

S7

S7

Time

Stream sequence 1

Calibration 1

Stream sequence 1

Assinning calibration 1

● In the case where a (semiautomatic) calibration/validation command is received during the

pause mode in the stream sequence status

Calibration (or validation) is immediately started.
When calibration (or validation) is completed for all the streams specifi ed for the selected

calibration (or validation) number, the operation mode changes to the stop mode in the
stream sequence status.

Pause

Stop

Stop

Run

W7

W7

W8

S8

S7

S7

Time

Stream sequence 1

Calibration 1

(Replaceable)

(Replaceable) Stream sequence 1

Assinning calibration 1 Calibration (or validation) start command

Calibration (or validation) stop command

● In the case where a (semiautomatic) calibration/validation command is received during the

run mode in the stream specifi cation status (for continuous measurement)

Calibration (or validation) is started when the measurement of the currently-measured

stream is completed.

Users at Level C or higher can cancel the calibration/validation command until the

measurement of the currently-measured stream is completed.

When calibration (or validation) is completed for all the streams specifi ed for the selected

calibration (or validation) number, the operation mode returns to the run mode in the stream
specifi cation status (for continuous measurement).

2nd Edition : May 11, 2012-00