25 simdis setup – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 349

Advertising
background image

<5. EtherLCD>

5-98

IM 11B08A01-01E

F2 (Main):

Display the User Programming screen(period script).

F6 (SYS #):

Specifi es an End of peak detection script number

• The operation accept user level above C+ only.
• When Delete is selected for User Program Delete, the relevant end of peak detection script

fi le is deleted. When a script is deleted, the auto start of the script is automatically changed
to NA. Any peak detection script in the status Enable cannot be deleted (an error message
is displayed).

• When the execution status is Enable and the user program is about to be deleted, the error

message “No execution because of executing” is displayed.

• Even when the user program is deleted, the script fi le in the SD card is not deleted.

However, the script on the user program is deleted.

Display Item

Name

Set Lower Limit Upper Limit

Unit

Remarks

SYS #
Delete User

Program

Do not delete

Delete

5.4.25 SimDis

Setup

1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5

SimDis Setup

GCM # 1
>

- Temperature unit deg C

- Det number of SIMDIS None
- Response factor updating Manual
- Section updating Auto
- Calib curve updating

Auto

Menu

Curve

GCM#

F1

F2

F3

F4

F5

F6

Figure 5.111

Example of SimDis Setup screen

F1 (Menu):

Displays the Table menu screen.

F4 (Convert): Displays the SimDis Calib Data Set screen.
F6 (GCM #): Specifi es a GCM number.

When setting an over limit value as a GCM number, the message “Over limit” is displayed in
the bottom line and the set value is not refl ected. For GCM set as NA for distillation on the GCM
setup (2/2) screen, “No distillation analysis” is displayed at the bottom line and setting is not
refl ected.
When the response factor updating is set to Yes, the response factor f on the peak setup-specifi c
screen of the cal run stream used by the set GCM is automatically updated.
When the section updating is set to Yes, the Section set start on the peak set-up specifi c screen
of the cal run stream used by the set GCM is automatically updated.
When the calib curve updating is set to Yes, the calib curve coef a and the calib curve coef b on
the SYS method set (2/4) screen are automatically updated.

• Det number of SIMDIS
Only detectors assigned to relevant GCM can be set up. On the GCM setup screen, the

SYS number belonging to each GCM is detectable and the detector number belonging to
each SYS on the SYS setup screen (2/5) can be detected.

2nd Edition : May 11, 2012-00