App.d-25, App.d, Start command (bm/be) run command is executed – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 470: Stop command (fm/fe/fi) stop command is executed

Advertising
background image

<Appendix D Action of External Input and Output Signals>

App.D-25

IM 11B08A01-01E

Calibration command (CM/CE/CA)

Calibration/Validation command is executed. Calibrations 4 to 6 and Validations 4 to 6 are not

supported.

1 2 3 4 5 6 7

For GC6

1 2 3 4 5

C M

CR LF

C A

CR LF

└┘

└┘

└┘

a

b

a

1 2 3 4 5 6 7 8 9

C E

CR LF

└┘

└─────┘

a

b

a:

Calibration/validation number (1 to 3 = Calibrations 1 to 3; 4 to 6 = Validations 1 to 3)

b:

Analyzer number (1 to 9 when M comes in the second digit, and 1 to 240 or 001 to 240 when E comes in the second digit)

Stream sequence change command (UM/UE)

Stream sequence change command is executed.

1 2 3 4 5 6 7

U M

CR LF

└┘

└┘

a

b

1 2 3 4 5 6 7 8 9

U E

CR LF

└┘

└─────┘

a

b

a:

Stream Sequence number (1 to 8)

b:

Analyzer number (1 to 9 when M comes in the second digit, and 1 to 240 or 001 to 240 when E comes in the second digit)

Start command (BM/BE)

Run command is executed.

1 2 3 4 5

For GC6

1 2 3 4

B M

CR LF

B E

CR LF

└┘

a

1 2 3 4 5 6 7

B E

CR LF

└─────┘

a

a:

Analyzer number (1 to 9 when M comes in the second digit, and 1 to 240 or 001 to 240 when E comes in the second digit)

Stop command (FM/FE/FI)

Stop command is executed.

1 2 3 4 5

For GC6

1 2 3 4

F M

CR LF

F I

CR LF

└┘

a

1 2 3 4 5 6 7

F E

CR LF

└─────┘

a

a:

Analyzer number (1 to 9 when M comes in the second digit, and 1 to 240 or 001 to 240 when E comes in the second digit)

2nd Edition : May 11, 2012-00

App.D