<7. Troubleshooting>

7-9

IM 11B08A01-01E

No.

MESSAGE

Alarm Contents

Threshold Item

Threshold

Threshold

Setting

482 REF PEAK ERR

Wrong reference

peak

Peak setting

Stream numbers

of reference peak

numbers are 1 to 31

and peak numbers

are other than 1 to

999.

Peak setting

483 CALC PEAK ERR

Peak calculation

error

N/A

484 TRACK PEAK ERR Peak tracking is not

possible

N/A

485 PEAK RECALC ERR Peak re-integration is

failure.

N/A

491 DET SIG OUT

Detector signal out of

range

AD conversion value

of detector

In the range of

0x800000 - 0x800003

or 0x7ffffc - 0x7fffff

N/A

493 COR ERR

Total compensation

error

Peak setting

Peak being executed

or peak being

processed does not

exist.

Peak setting

497 BL CHRMT ERR

Baseline

chromatogram error

N/A

498 CHRMT SAVE ERR Saving

chromatogram fi le

failure

SD card

Error to write SD

N/A

521 OVEN1 POWER ON OVEN1 power on

Power supply

Power supply turned

on

N/A

522 OVEN2 POWER ON OVEN2 power on
523 OVEN3 POWER ON OVEN3 power on
551 PRG SEND ERR

Program send error File transfer

Error to transfer fi le

N/A

552 COMMON INFO

ERR

Common information

error

File transfer

Error to transfer fi le

N/A

553 CARD START ERR Card start error

System setting

Mismatch between

slot setting and card

ID.

Slot setting screen

554 CARD COM ERR

Card communication

error

N/A

556 COM SYS MSG

System error

message

N/A

557 SCRIPT SEND ERR Script send error

File transfer

Error to transfer fi le

N/A

558 SCRIPT EXE ERR

Script execution error File transfer

Error to transfer fi le

N/A

559 SCRIPT DEL ERR

Script delete error

File transfer

Error to transfer fi le

N/A

560 RESET MSG

Reset message

Power supply

Power supply turned

on

N/A

561 SCRIPT TIMEOUT

ERR

Script execution

time-out error

Script execution time Script execution is

not completed within

3 seconds

Script File

562 MODBUS CLIENT

ERR

MODBUS CLIENT

overlap error

IP address,

equipment ID

ID address or

equipment ID is

overwrapped.

IP address,

equipment ID

*:

Cases resulting in the mismatch of setting parameters

1: The ratio between the SYS analysis cycle and GCM main cycle is not n: 8 (n = 1 to 8).

2: The pause time is not set at the last SYS analysis cycle in the GCM main cycle.

3: Gate ON/OFF times for peaks to be detected by each detector overlap. For the zone gate, the gate OFF time for a peak must

be the ON time for the next peak to be detected in the same zone.

4: The SYS analysis cycle and the GCM main cycle do not match when a programmed temperature oven is mounted.

5: The sampling rates of the detectors belonging to the same SYS differ (different sampling rates can be set for the detectors

belonging to the same GCM).

6: The peak detection completion time does not fall within the range from 5 seconds to the analysis cycle.

7: There is no stream in the current stream sequence number when the measurement state is set to the stream sequence.

8: All the SYSs belonging to GCM are set to “not used.”

9: The gate ON/OFF time does not fall within the range of the SYS analysis cycle.

10: No stream belongs to GCM.

11: During the calorifi c calculation, the units of each peak in streams do not match, or units of wt, mol or user-defi ned ones

(including unused ones) are used.

12: Among multiple SYSs, there is at least one SYS whose peak number is zero.

13: The peak assignment number or peak number is zero (all peaks are set to “not used”).

14: The peak process is set to the base, signal or noise level, and the level action time falls in the time between gate ON and

OFF.

15: Data in the main cycle, preparation time, or pause time is invalid.

16: The action time at the signal, base, or noise level does not fall within the range of the SYS analysis cycle.

2nd Edition : May 11, 2012-00

7