<7. Troubleshooting>

7-12

IM 11B08A01-01E

7.2.3 Malfunctioning

Chromatogram

It is available to use three types of detector such as TCD, FID and FPD. Troubleshooting diagram
for malfunctioning chromatogram is different according to the shown as followings.

(1) TCD

Troubleshooting diagram for malfunctioning chromatogram is shown as followings.

F0304.ai

Malfunctioning chromatogram

Bridge balance

of detector is good

Yes

Yes

No

Yes

No

Yes

No

Yes

No

Yes

No

Yes

No

Yes

No

Yes

No

Yes

No

No

Yes

No

Yes

No

Turn on detector

DET CURRENT

ERROR is occured

Status of

detector is ON

Replace detector

Wiring for

detector is good

Replace wiring for detector

Replace detector board

Adjust current with volume

on detector board

Current value

is as specified in operation

data

Current can be adjusted

End

Baseline of

detector is malfunctioning

Column system failure or

sampling system failure

Temperature

of oven is stable

Wait for stabilization of

oven temperature

External piping is clean

Use clean piping

Bridge

balance of detector is

good

Repalce detector

Insulation

resistance of detector is

good

Replace wiring for detector

Wiring for detector is

good

Replace detector board

Figure 7.4

Troubleshooting Diagram for Malfunctioning Chromatogram (TCD)