App.e-18 – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 491

Advertising
background image

<Appendix E Computation Scheme>

App.E-18

IM 11B08A01-01E

(b) ASTM D3588-81

The GC8000 converts each peak result obtained in volume percent into the one in mole fraction,

and calculates the real gas calorifi c value, real gas specifi c gravity, compressibility factor, and

Wobbe index of the sample gas from the ideal gas calorifi c value, ideal gas specifi c gravity,

compressibility factor, and compressibility summation factor of each component.

<Formula>

Z = 1 – (

)

2

+ 0.005(2x

H

– x

H

2

)

x

i

H

r

: real gas calorific value (total or net)

H

: ideal gas calorific value (total or net)

H

i

: ideal gas calorific value of component i (total or net)

G

r

: real gas specific gravity

G

: ideal gas specific gravity

G

i

: ideal gas specific gravity of component i

z

: compressibility factor

z

i

: compressibility factor of component i

b

i

: compressibility summation factor of component i

x

i

: mole fraction of component i

x

H

: mole fraction of hydrogen

x

Vi

: volume percent of component i

W : Wobbe Index

H

r

=

z

H

G

r

=

z

0.99959G

W =

G

r

H

r

H =

x

i

H

i

G =

x

i

G

i

x

i

=

x

Vi

z

i

x

Vi

z

i

b

i

Where:

2nd Edition : May 11, 2012-00

Advertising