background image

<5. EtherLCD>

5-69

IM 11B08A01-01E

Table 5.3

Synchronization rules table

Destination

Source

Measurement Cal

Val

Calibration

run

Sample run

Measurement

Ч

Ч

Ч

Cal

Ч

Ч

Ч

Val

Ч

Ч

Ч

Calibration run

Ч

Ч

Ч

Ч

Ч

Sample run

Ч

Ч

Ч

Ч

○, ▲: Synchronization possible, see Synchronize items for synchronization items.
*:

The

▲ symbol in the Synchronization Rules Table signifi es (Cal →Measurement, Cal →Val) and he synchronization items ▲

(process assignment, reference peak, etc.) cannot be synchronized.

×:

Cannot be synchronized.

The following is an example of setting error.

Synchronized destination: Stream 2ーPeak 1

Stream 1ーPeak 2

F0502.ai

Stream 1ーPeak 1

Indirect synchronization

Contradicts the rule “Cannot set peak within the same stream to synchronized destination.”

2nd Edition : May 11, 2012-00