<Appendix D Action of External Input and Output Signals>

App.D-3

IM 11B08A01-01E

Table 2

Example of Analog Output Settings

Output stream #1 Output peak #1 Output stream #2 Output peak #2

A/O1

1

1

None

None

A/O2

1

2

None

None

A/O3

1

3

None

None

A/O4

1

4

None

None

A/O5

2

1

None

None

A/O6

2

2

None

None

A/O7

2

3

None

None

A/O8

2

4

None

None

F0501.ai

*1

*1

*1

*1

Stream 1

End of Analysis

Stream 2

End of Analysis

Stream 1

End of Analysis

Stream 2

End of Analysis

AO1

AO2

AO3

AO4

AO5

AO6

AO7

AO8

Analog Hold Output

End of Analysis: Time when the final peak detection is completed or peak detection stop time.
*1:

One to two seconds (depending on analysis specifications)

End of Analysis:

End of peak detection time in SYS method

*1:

One to two seconds (depending on analysis specifi cations)

Figure 3 Example of Actions of Analog Output

When Actual Stream is Set for Output Stream Number (with Stream

Identifi cation Signal)

Select Normal output for the AO chromatogram on the A/O setting screen to open the analog

hold output (analysis result output) setting screen (Figure 1). Specify the GCM number, output

stream number, peak number, and the upper and lower limits of the partial range for the peak of

which analysis results are to be output.

Table 3 and Figure 4 show the settings for one GCM with six streams where a stream

identifi cation signal is used.
Table 4 and Figure 5 show the settings for two GCMs with three streams each where a stream

identifi cation signal is used.