Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 418

Advertising
background image

<7. Troubleshooting>

7-5

IM 11B08A01-01E

No.

MESSAGE

Alarm Contents

Threshold Item

Threshold

Threshold

Setting

90 TMP1 SENS C S/C Temp. sensor 1 short

circuit

Temperature sensor

value

Temperature falls

below the lower limit

of –20°C

N/A

91 TMP2 SENS C S/C Temp. sensor 2 short

circuit

92 TMP3 SENS C S/C Temp. sensor 3 short

circuit

93 TMP1 SENS P S/C

Overtemp protect

zone 1 short circuit

N/A

94 TMP2 SENS P S/C

Overtemp protect

zone 2 short circuit

95 TMP3 SENS P S/C

Overtemp protect

zone 3 short circuit

111 ELEC PRESS

DOWN

Low electronic purge

pressure

N/A

112 OVEN1 PRESS

DOWN

Low oven 1 pressure

N/A

113 OVEN2 PRESS

DOWN

Low oven 2 pressure

114 OVEN3 PRESS

DOWN

Low oven 3 pressure

115 PURGING1

Oven 1 purging

N/A

116 PURGING2

Oven 2 purging

117 PURGING3

Oven 3 purging

121 DET1-1 A/D ERR

Detector 1-1 A/D cal.

error

Detector card value Exceed internal

upper and lower limit

values

N/A

122 DET1-2 A/D ERR

Detector 1-2 A/D cal.

error

123 DET2-1 A/D ERR

Detector 2-1 A/D cal.

error

124 DET2-2 A/D ERR

Detector 2-2 A/D cal.

error

125 DET3-1 A/D ERR

Detector 3-1 A/D cal.

error

126 DET3-2 A/D ERR

Detector 3-2 A/D cal.

error

127 DET1-1 FLAME

OUT

Detector 1-1 A/D

fl ame out

FID & FPD fl ame

detector level

Flame detectors

for FID & FPD drop

below detection level

Detector signal

setting screen

128 DET1-2 FLAME

OUT

Detector 1-2 A/D

fl ame out

129 DET2-1 FLAME

OUT

Detector 2-1 A/D

fl ame out

130 DET2-2 FLAME

OUT

Detector 2-2 A/D

fl ame out

131 DET3-1 FLAME

OUT

Detector 3-1 A/D

fl ame out

132 DET3-2 FLAME

OUT

Detector 3-2 A/D

fl ame out

133 DET1-1 CURRENT

ERR

Det 1-1 current out-

of-range

TCD current value

TCD current value

exceeds 350 mA for

H2-He mix, and 150

mA for others.

Current applied to

the detector drops

below 60 mA during

Run mode.

N/A

134 DET1-2 CURRENT

ERR

Det 1-2 current out-

of-range

135 DET2-1 CURRENT

ERR

Det 2-1 current out-

of-range

136 DET2-2 CURRENT

ERR

Det 2-2 current out-

of-range

137 DET3-1 CURRENT

ERR

Det 3-1 current out-

of-range

138 DET3-2 CURRENT

ERR

Det 3-2 current out-

of-range

141 EPC1 COM ERR

EPC 1

communication error

EPC communication EPC communication

fails 5 consecutive

times

N/A

142 EPC2 COM ERR

EPC 2

communication error

143 EPC3 COM ERR

EPC 3

communication error

2nd Edition : May 11, 2012-00

7

Advertising