<1. Overview>

1-70

IM 11B08A01-01E

Pause

Stop

Run

W7

W7

W8

S8

S7

S7

Time

Stream sequence 1

Calibration 1

Stream sequence 1

Start of automatic calibration

● In the case where automatic calibration starts during the run mode in the stream

specifi cation status (for continuous measurement)

Calibration (or validation) is started when the measurement of the currently-measured

stream is completed.

When all the calibration (or validation) patterns of which automatic execution is enabled are

completed, the operation mode returns to the run mode in the stream specifi cation status
(for continuous measurement).

Run

W7

W7

W8

W1

S1

S1

S8

W1

W1

S1

S7

S7

Time

Stream specification 1 (Continuous) Calibration 1

Stream specification 1 (Continuous)

Start of automatic calibration

● In the case where automatic calibration starts during the run mode in the stream

specifi cation status (for N times of measurement)

Calibration (or validation) is started when the Nth round of the measurement of the currently-

measured stream is completed.

When all the calibration (or validation) patterns of which automatic execution is enabled are

completed, the operation mode changes to the stop mode in the stream specifi cation status
(for N times of measurement).

Run

W7

W7

W8

S8

W1

S1

W1

S1

S7

S7

Time

Stream specification 1 (Twice)

Calibration 1

Stream specification 1 (Twice)

Start of automatic calibration

Run command

Stop

Stop

● In the case where automatic calibration starts during the stop mode in the stream

specifi cation status

Calibration (or validation) is immediately started.
When all the calibration (or validation) patterns of which automatic execution is enabled are

completed, the operation mode returns to the stop mode in the stream specifi cation status.

Run

W7

W7

W8

S8

S7

S7

Time

Stream specification 1

Calibration 1

Stream specification 1

Start of automatic calibration

Stop

Stop

● In the case where automatic calibration starts during the pause mode in the stream

specifi cation status

Calibration (or validation) is immediately started.