Data to be received:

Operation requests (stream sequence

setting, stream setting, run, stop,

pause, and range change)

External I/O Cutoff Output:

Number of outputs: 1 or 2

Function: Monitoring the purge air pressure

in the electronics section,

applying power (24 V DC) to the

communication converter/the signal

interrupter when the state is normal.

Signal interrupter (Rack-mounted type:

K9806AE):

FM-X, ATEX, IECEx, TIIS:

Additionally

required.

RS-422 is outputed signal is

interrupted by power supply OFF

signals from the external I/O cutoff

output.

FM-Y: Not

required

Note: For installation in hazardous area, an explosionproof

structure authenticated by a relevant certifi ed body
shall be prepared by customer.

Communication converter (Rack-mounted type:

K9806AS, Desktop type: K9806AT):

RS-422/232C

communication

converter with signal interrupter

function.

Signal is interrupted by power supply

OFF signals from the external I/O

cutoff output.

Note: Two units are required for 2 channel communication.
Note: For installation in hazardous area, an explosionproof

structure authenticated by a relevant certifi ed body
shall be prepared by customer.

4.1.3

Connection to GCCU MarkII (GC Computing Unit)

Both Ethernet and Serial communication are

available.

Note: Applicable only to 1GCM
Note: Analog hold output in the analog output function

cannot coexist with the GCCU communication.

(1) Ethernet communication

The same port as analyzer network is used.

Refer to 4.1.1 Connection to Analyzer network.

Communication standard:

Ethernet

Protocol: Modbus over TCP

Communication speed:

100Mbps

Note: SHDSL interface (K9802PB) is provided as an

option (1 channel only).

(2) Serial communication (Option)

Channel: 1

Communication standard:

RS-422

Transmission: Full duplex (4-wire system)

Protocol: Modbus

Start-stop (asynchronous) communication:

Start bit 1, Data bits 7, Parity bit 1, Stop

bit 1

Parity check: Even

Communication speed:

9600 bps

Data format: ASCII

Communication control:

None

Terminal: Terminal connection (Phoenix terminal)

External I/O Cutoff Output:

Number of outputs: 1

Function: Monitoring the purge air pressure

in the electronics section,

applying power (24 V DC) to the

communication converter when the

state is normal.

Communication converter (Rack-mounted type:

K9806AS, Desktop type: K9806AT):

RS-422/232C

communication

converter with signal interrupter

function.

Signal is interrupted by power supply

OFF signals from the external I/O

cutoff output.

Note: For installation in hazardous area, an explosionproof

structure authenticated by a relevant certifi ed body
shall be prepared by customer.

4.2 Input/Output

System alarm 1 and Annunciator are provided as

standard contact outputs. Other Input/Output can

be added depending on specifi cations.

Note: Up to 5 optional cards are selectable.
Note: The maximum number of the contact inputs is 32

and outputs is 20, with any combination of contact
input cards, contact output cards, and contact I/O
cards.

4.2.1

Contact Output for System Alarm 1

Number of contact outputs: 1

Function: Starting operation when System alarm

1 occurs.

Contact specifi cation:

Relay contact output, c-contact (NC/

NO/COM)

Contact rating:

30 V DC, 100 mA

Contact operation: Open/Close

Contact

System alarm

Between NC and

COM terminals

Between NO and

COM terminals

On operation

Open

Close

Off operation

Close

Open

Power supply: OFF

Open

Close

4.2.2

Contact Output for Annunciator

Number of contact outputs: 1

Function: Outputting the state of the purge air

pressure in the isothermal oven and

the electronics section (State 1 or

State 2)

State 1: Outputting the states of override

mode ON, power supply of the

analyzer OFF, purging, or insuffi cient

pressure

State 2: Outputting the state after the purge

Contact specifi cation:

Relay contact output, c-contact (NC/

NO/COM)

Contact rating:

30 V DC, 100 mA