<5. EtherLCD>

5-49

IM 11B08A01-01E

F1 (Menu):

Displays the Table Menu screen.

F5 (Oven #): Sets an oven number.
F6 (Det #):

Sets a detector number.

TIP

• The oven numbers set to Not provided on the Temperature Control Set cannot be selected.
• If the Detector type has been set to None on the Detector Set screen, all the options are shown as *

(asterisk).

• The initial screen displays Oven number 1 and Det number 1.
• When A/O chromatogram is set to None, items for Auto gain and Whole gain value are

displayed but an asterisk is given for set value.

• The initial screen displays the smallest oven number among those set in the isothermal

ovens on the Temperature Control Set.

• The initial screen displays Oven number 1 and Det number 1.

Display Item

Name

Set Lower Limit Upper Limit

Unit

Remarks

Oven #

Oven number

1

3

Det #

Detector number

1

2

Detector type

0) None

1) TCD

2) FID

3) FID-MC

4) FPD

Auto gain

0) None (default)

1) Individual gain

2) Whole gain *1

*1 Only for the A/O

chromatogram

Whole gain

value

0

15

0(default)

*1 Only for the A/O

chromatogram

Auto zero

0)Automatic feed (default)

1)Auto zero value

2)Disabled

* If the value is Automatic feed,

the value is set to the zero value

automatically obtained at the

auto zero time.

* If the value is Auto zero value,

the value is set to the auto zero

value at the start of the analysis.

Auto zero

value

−1000.0000 1000.0000

Former chromatogram standard

value

Auto zero

time

0

21600

If the value is 0.0 seconds, the

value is set to the zero value

automatically obtained at the

start of the analysis cycle.

* Upper Limit will use the

shortest analysis cycle for the

SYS analysis cycle that the

detector belongs to.

2nd Edition : May 11, 2012-00