<Appendix C Standard Specifi cations>

App.C-2

IM 11B08A01-01E

ATEX:

II2G Ex d px IIB+H

2

T1...T4 Gb

IECEx: Ex d px IIB+H

2

T1...T4 Gb

TIIS:

Ex pd IIB+H

2

T1 to T4

EMC standard: EN61326-1 Class A (Emission)

EN61326-1, EN61326-2-3 (Immunity)

Korea Electromagnetic Conformity

Standard

Note: ATEX, IECEx and TIIS (/KC) only

Safety standard:

ATEX, IEXEx: EN61010-1

FM: FM3810

(IEC61010-1)

CSA: CSA61010-1

Protection degree of enclosure:

NEMA3R, Equivalent to IP54 (dust and

water resistant structure)

Display: LCD (or without display) and LED

(POWER/ALARM/RUN)

Operating display unit:

Touch panel (or

without operating display unit)

Coating: Polyurethane baked fi nish

Main body: Silver gray (Munsell 3.2PB 7.4/1.2

or its equivalent)

Gauge: Mint green (Munsell 5.6BG 3.3/2.9 or

its equivalent)

Weight:

Wall-mounting version Self-standing version

Type 1

approx. 100 kg

approx. 140 kg

Type 2

approx. 155 kg

approx. 190 kg

Type 3

approx. 200 kg

approx. 220 kg

Other functions:

• A real-time clock with back-up

batteries is mounted in the control unit

(except for the TIIS specifi cation).

• Data storage

The GC8000 can save 1 week of

chromatogram data, any chromatogram

data (up to 20 per GCM) 30 days of

analysis results, and 100 calibration

factors. The number of saved data

depends on analysis conditions. Saved

data can be read and displayed on both

the HMI and the PC.

• The GC8000 has a function for

limiting the supply of hydrogen when

the FID/FPD fl ame is extinguished.

2. Element

Specifi cations

2.1 Isothermal Oven

Volume:

Large isothermal oven:

Approximately 45 L

Standard isothermal oven: Approximately 31 L

Oven temperature range:

55 to 225°C (Temperature can be set

in one-degree step.)

Temperature stability: ±0.03°C

Temperature control accuracy:

±0.03°C

Temperature control: PID

Temperature sensor: Pt100 Ω RTD

Other functions:

Over-heating prevention function

2.2 Liquid-sample Valve with Vaporizer

Sample pressure:

0 to 3 MPa

Sample temperature: 150°C or lower

Sample volume:

0.25, 0.5, 1, 2, and 3 μL

Vaporizing section:

LSV temperature range:

Oven temperature +5 to 250°C

Temperature stability:

±1°C

Temperature control accuracy:

±1°C

Temperature control: PID

Temperature sensor:

Pt100 Ω RTD

Other functions:

Over-heating prevention function

3. Utility

3.1 Power

Power supply:

100/110/115/120/200/220/

230/240 V AC ±10%, 50/60 Hz ±5%

Note: Type 3 requires 200 V or more.

Protection devices, such as a breaker, are required
to avoid overcurrent.

Wiring method:

FM:

Conduit wiring (3/4NPT(F))

ATEX, IECEx:

Cable packing (G3/4(F),

3/4NPT(F), M25x1.5(F))

TIIS:

Cable packing (G3/4(F), 3/4NPT(F))

Note: Cable packing for TIIS is provided by Yokogawa.

Other conduit wiring or cable packing should be
prepared by the user.

Wiring connection:

FM-X, ATEX, IECEx, TIIS:

Explosion proof enclosure

FM-Y: Control

unit

Maximum rated power::

Type 1: 0.8 to 1.6 kVA

Type 2: 1.4 to 2.9 kVA

Type 3: 2.0 to 4.3 kVA

3.2 Utility gas

Note: It may vary depending on application.

3.2.1 Instrument

air

Pressure: 350 to 900 kPa

Flowrate:

Type 1:

100 to 140 L/min

Type1 with FPD: 130 to 200 L/min

Type 2:

150 to 210 L/min

Type2 with FPD: 180 to 270 L/min

Type 3:

200 to 280 L/min

Temperature: –10 to 50°C

Dew point:

–20°C or lower (condensation of

compressed air must be avoided at the

ambient temperature.)

Oil:

5 ppm or less

Cleanliness: Must be free from dust, corrosive

elements, and toxic elements.

Connection: Rc1/4 or 1/4NPT (F)

3.2.2

Carrier gas, combustion gas for FID/FPD,

make-up gas for FID/FPD

Types: H

2

, N

2

, He, or Ar

2nd Edition : May 11, 2012-00